Water bath


Water bath


Purpose of water bath : Water bath is required to incubate  bottles of culture media, liquid in flasks or other large containers and when incubating samples in test tube racks.


Specification of water bath for the district laboratory:


  1.It is unstirred with hydraulic thermostat or electronic temperature control.

  2.It can operate temperature range from ambient to 60°C or above with safety .

  3.Having a 4 –litter capacity,

  4.Fitted with a thermometer.

  5.Supplied with a lid.



Use and care of water bath:


  1.Read the manufacture’s instruction , prepare a stock record care and written SOPs converging the use , care maintenance of water bath.

  2.Fill the bath and maintain its level with distilled water or boiled filtered rain- water.

  3.To minimize the growth of the microorganism in the water , use a bactericidal agent such as merthiolet at a dilution of 1 in 10000

  4.Before incubating check the temperature of the water.

  5.Ensure that the level of the water is above the level of whatever is being incubated.

  6.Use the lid to prevent loss of heat from the bath and to minimize particles from entering the water.

  7.Clean the water bath regularly , taking care, not to damage the heating unit .

  8.Unplug the bath when not using .

  9.Check the bath for corrosion the flex for wear and the plug for any loose connection for every 3-6 months.